User Contributed MET/CAL Procedure ============================================================================= INSTRUMENT: EMC 11 Frequency Sub DATE: 21-Dec-95 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 7 NUMBER OF LINES: 60 CONFIGURATION: HP 3325A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 IEEE [TERM LF] 1.002 HEAD {} 1.003 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.004 HEAD {º Frequency Accuracy º} 1.005 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.006 HEAD ÍÍ Frequency Accuracy ÍÍ 1.007 JMP 2.001 1.008 EVAL 2.001 IEEE [@3325]FU1;AM 100MR;FR 100HZ 2.002 IEEE RES:1[13][10] 2.003 IEEE GTL[13][10] 2.004 DISP Manually tune the UUT to .1 kHz and peak the meter. 2.005 IEEE FR[13][10][I] 2.006 ACC 100H 0.01% 2.007 MEME 2.008 MEMC H 5.0U 3.001 IEEE [@3325]FU1;AM 100MR;FR 500HZ 3.002 IEEE GTL[13][10] 3.003 DISP Manually tune the UUT to .5 kHz and peak the meter. 3.004 IEEE FR[13][10][I] 3.005 ACC 500H 0.01% 3.006 MEME 3.007 MEMC H 5.0U 4.001 IEEE [@3325]FU1;AM 100MR;FR 2.5KH 4.002 IEEE GTL[13][10] 4.003 DISP Manually tune the UUT to 2.5kHz and peak the meter. 4.004 IEEE FR[13][10][I] 4.005 MEM/ 1000 4.006 ACC 2.500kH 0.01% 4.007 MEME 4.008 MEMC kH 0.10% 5.001 IEEE [@3325]FU1;AM 100MR;FR 10KH 5.002 IEEE GTL[13][10] 5.003 DISP Manually tune the UUT to 10 kHz and peak the meter. 5.004 IEEE FR[13][10][I] 5.005 MEM/ 1000 5.006 ACC 10.000kH 0.01% 5.007 MEME 5.008 MEMC kH 0.10% 6.001 IEEE [@3325]FU1;AM 100MR;FR 25KH 6.002 IEEE GTL[13][10] 6.003 DISP Manually tune the UUT to 25 kHz and peak the meter. 6.004 IEEE FR[13][10][I] 6.005 MEM/ 1000 6.006 ACC 25.000kH 0.01% 6.007 MEME 6.008 MEMC kH 0.10% 7.001 END